Conductivity
Conductivity Meter w/ Relay Outputs
The LF1010 Conductivity Meter has been designed for the measurement of conductivity, using standard 2 electrode or 4 electrode conductivity cells. When used in connection with 4-electrode cell it provides high accuracy over a large range of conductivity. For measurements of ultra pure water it should be used with a two electrode cell with a low cell constant 0.1) for a high accuracy and sensitivity. The meter features a rugged packaging and relay outputs, choice of power supply, and temperature monitoring via Pt100 or Pt1000 platinum resistance thermometers. FDA compliant.
- Broad measurement range (with 4 electrode cells)
- Works with 2 or 4 electrode cells
- Temperature Compensation (with Pt100 or Pt1000)
- Relay Alarm outputs
- Min Max Peak Hold (option )
- AC or DC powered
- FDA and USP645 compliant sensor cells
Key Specifications
Supply Voltage
230VAC or 115VAC or 24VAC or 24VDC
Conductivity Probe Type
2 or 4 pole, cell constant 0.080 to 9.999
Conductivity Range
0...2.000 uS/cm | 0...2,000uS/cm | 0...200 mS/cm | 0...2,000mS/cm
Accuracy
0.5% of the measuring range
Temperature Compensation
pt100 or pt1000 | non linear for ultra pure water and natural water or linear programmable from 0.000..9.999 %/K
Temperature Compensation Range
-50 ... 200.0C
Outputs
Two relays: SPDT < 250 V AC < 250 VA < 2 A, < 300 V DC < 50 W < 2 A
CE Conformity
EN 61326-1:2013 EN 60664-1:2007
Power Consumption
Max 3.5VA
Available Accessories
Available Accessories
Calibration and service is available for this item.