The standard procedure to separate particles from a solutions is a gravimetric filtration. This method is time consuming and can result in small amounts of evaporation of the liquid. In the sugar laboratory the automatic filtration systems AutoFilt reduces the filtration time by at least a factor of 10 and ensures a turbidity free solution in a very short time. The AutoFilt filtration system is specially designed to speed up
sample preparations for the polarimetric analysis of sugar solutions
using Near Infra Red polarimeters (Saccharimeters).
The AutoFilt obviates the need to clarify sugar and avoids the use of
lead acetate and aluminum sulfate.
The unit comprises a sample compartment, with filter holder, a pump,
and a stand to hold the sample chamber and collect the filtrate. The
unit operates with compressed air which is provided by a small
compressor (or central air). The compressed air forces the liquid in
the sample holder through a filter and into the filtrate collection
flask. The operation is automated with the push of a button
Filtration Pressure | max. 5 bar - 0.5 Mpa |
Sample Volume | max. 250 ml |
Measurement Volume | 50 ml sensor controlled |
Operating temperature (environment) | +10 - +40 °C |
Wavelength section | Via touch screen |
Power required | 100V-240 VAC, 47-63 Hz, 40 W |
Pressurized Air supply | max. 10 bar - 1,0 Mpa oil free |
Air Supply tube | 6 mm OD |
Waste water tube | 16 mm ID |
Filter Area | approx. 78.5 cm sq |
Filtration Pressure | 6 Bar |
Filtration Volume | 250 ml |
Operator Interface | Touch screen for direct functions and alphanumeric inputs |
Part Number | Model | Pice guide |
TID 00501 | AutoFilt (Compressor not included) including 100 150mm filter pads | Contact us for more information or a quotation |
TID 03709 | Compressor (optional) up to 10 Bar wiht pressure control valve | |
TID 12128 | Filter pads 150 mm diam |