AM 215A Digital Comparator - Voltage/Current Meter

Features
  • Compact DIN size
  • Two-stage alarm setting
  • RoHS compatible product
  • Dimensions: 48 (w) x 24 (H) x 87.8 (D) mm
  • Display: Full 4-digit LED (red) 8 mm
  • Input: DC voltage/current
  • Measurement/input/scaling range: ±99.99 mV, ±999.9 mV,
    ±9.999 V, 1 to 5 V, 4 to 20 mA
  • Sampling speed/response speed: Max. 25 counts/second
  • Power supply: 24 VDC
  • Output: Relays, Optional Analog, RS485

Measurement Ranges DC Voltage

Model Range Display
adjustable
Input Impedance Maximum
Allowed Input
Accuracy
AM 215A-11 ± 99.99 mV
offset ± 9999
Full scale
0 to ± 9999
100 MOhm or more ±50V ±(0.03% of reading +2 digits)
At 23°C ±5°C
AM 215A-12 ±999.9 mV
AM 215A-13 ±9.999 V ~ 1 MOhm
AM 215A-1V 1 - 5V

Measurement Ranges DC Current

Model Range Display
adjustable
Input Impedance Maximum Allowed Input Accuracy
AM 215A-2A 4-20 mA
offset ± 9999
Full scale
0 to ± 9999
~ 50 Ohm ±50mA ±(0.1% of reading +2 digits)
At 23°C ±5°C

Common Specifications

Measurement function Select either DC voltage or DC current (single range)
Input circuit Single ended<
Operation type Double integral
Sampling speed Maximum per second at 50 Hz, or 15 per second at 60 Hz
Display Red 7-segment LED display (character height: approximately 8 mm)
Polarity display A minus sign is displayed if the operation result is negative.
Out-of-range warning "oL" or "-oL" is displayed with respect to the input signal when the value is outside the display range.
Maximum display ±9999 (full 4 digits)
Decimal point Can be set at any position using the front-panel flat switches
Zero display Leading zero suppress
Backup Settings are held in EEPROM (guaranteed for 100,000 writes)
Operating temperature and humidity ranges 0 to 50°C, 35 to 85% RH (no condensation)
Storage temperature and humidity ranges -20 to 70°C, 60% or lower RH (no condensation)
Supply voltage DC 24 V ±20%
Consumed power Approximately 4 W
Relay Output ratings DC24V 1A
Photocoupler Output Max 30V, 50mA - output saturation 1.2V max at 50 mA
External dimensions 48 x 24 x 87.8mm
Weight 100 g
Withstand voltage DC 500 V for one minute across power terminals/input
terminals and each output terminal
DC 500 V for one minute across input terminals/each output terminal
AC 1500 V for one minute across case/power terminals, input terminals, and each output terminal
Insulating resistance 100 M ohm or higher with DC 500 V across the terminals listed above

Mode Settings

Setting ConditionHI GO LO HI Limit > LO Limit
HH HI GO HH Limit > HI Limit
GO LO LL LO Limit > LL Limit
Hysteresis All limit values can be set from 1 to 999

Comparator Conditions

HI/GO/LO Mode   Output
Comparator conditions AL1 AL2 Out1 Out2 Out3
Measurement value > HI limit value off on off off on
LO limit value ≤ Measurement value ≤ HI limit value off off off on off
LO limit value > Measurement value off off on off off
HH/HI/GO Mode   Output
Comparator conditions AL1 AL2 Out1 Out2 Out3
Measurement value > HH limit value on on off on on
Measurement value > HI limit value on off off on off
HI limit value ≥ Measurement value off off on off off
GO/LO/LL Mode   Output
Comparator conditions AL1 AL2 Out1 Out2 Out3
Measurement value ≥ LO limit value off off off off on
LO limit value > Measurement value off on off on off
LL limit value > Measurement value on on on on off

Analog Output Option

Output function Set to 50 Hz or 60 Hz, depending on the power supply frequency being used.
Output specifications Mode Load Resistance Accuracy Ripple
4-20 mA 0-250 ohm ±(0.5% of full scale) 25 mVp-p max
0-10 V 10 k ohm min ±(0.5% of full scale) 50 mVp-p max.

Digital RS485 Output Option

Synchronization method Start-stop synchronization
Communication method 2-wire half-duplex (polling selecting)
Transfer rates 19200, 9600, 4800, 2400 bps
Start bit 1 bit
Data length 7 bits
Error check Even parity, block check character (BCC) checksum
Stop bits 2 bits
Character code ASCII code
Delimiter CR + LF
Transmission control
protocol
None
Used signal name Non-inverted (+), inverted (-)
Maximum connected units Maximum 31 meters
Maximum line length 500 meters (total)
 

Ordering Information

AM-215A- powered by 24V DC select range 11,12,13,1V 2A above
Select comparator output -1 relay -2 Photocoupler
$258
AM-215A- with Analog or RS-485 output
Select optional outputs
-1 none
-4 RS 485
-6 4-20 mA
-7 0-10 V
$273

This is a world wide web document page design layout and HTML code copyright ©2002-2016 by Topac Inc. - ALL RIGHTS RESERVED