You can purchase online where you see the check mark |
![]() |
Click on the assistant to view the contents of your shopping cart |
Digital Panel Meters Set Point Comparator |
![]() |
Digital panel meters from Asahi Keiki offer impressive specifications, reliability and low cost for a variety of applications |
AM-749 | Features
|
![]() |
DC Voltage Measurement |
Accuracy +/- 0.03% of reading +/- 2 digit (23 C +/-5 C |
DC Current Measurement |
Accuracy +/- 0.1% of
reading +/- 2 digit (23 C +/-5 C +/- 0.3% of reading +/- 1 digit only for 25 range |
General
Specifications |
Measurement Block | |
Measurement function | Specify one measuring range from DC voltage, current ranges below. |
Input Circuit | Single Ended |
Operation method | dual slope |
Input Bias Current | 100 pA |
Conversion Rate | 12.5 /sec (50 Hx) 15 /sec (60Hz) |
Normal mode rejection | > 50dB (50/60Hz) |
Display | 7 segment LED, 10 mm red monitor sub display LED 8mm green |
Polarity | "-" automatic when result is negative |
Decimal point | Set to any position |
Over-range indication | 'OL' LED flashes |
Maximum Display | +/- 99999 ( 5 digits ) |
Decimal point | set to any position |
Zero Display | Reading "zero" suppression |
External Control | |
Hold | Shorting between COM and S/H terminal or "0" level |
Start | Open between COM and S/H terminal or "1" level |
Digital zero | Value just before COM and DZ terminals are shorted, or input is set to "0", is displayed as "zero" and that value is stored |
Peak Hold | Respective function is activatedd when COM and PH terminals are shorted or input level is set to "0" |
Valley Hold | |
Peak Valley Hold | |
Pattern selection | Anyone of eight
patterns is settable in combination of COM terminal with any of P SEL terminals
0, 1 and 2 "0" level less than 5 volts "1" level 3 - 30 volts input current less than -5 mA |
Comparison Block | |||||||||||||||||
Control Method | Computation by microcomputer | ||||||||||||||||
Measuring range comparison |
high/Low limit setting including polarity, -99999 to + 99999 | ||||||||||||||||
Setting condition | By
sampling period No setting condition for set value |
||||||||||||||||
|
|||||||||||||||||
Comparison
Relay contact capacity |
250 Volts AC
0.2A resistive load 120 Volts AC 0.5A resistive load 28 Volts AC 1 A resistive load |
||||||||||||||||
Photo-coupler output | sink current 20
mA max )less than 30 V. Output saturation voltage < 1.2 Volts at 20 mA |
||||||||||||||||
Hysteresis | Settable from 1 to 9999 digits for each comparison set value | ||||||||||||||||
External Control | Comparison
operation is suspended when COM and r1R.RES terminals are shorted or iput
level is set to "0" "0" level less than 1.5 V "1" level 3 to 30 V - input current < - 5mA |
Common Specifications | |
Memory Backup | Set data stored for 10 years by ERPROM |
Temperature and humidity | 0 - 50 degs C, 35% to 85% RH |
Power Supply | AC 90 - 264
V |
Operation temperature | 0~ 50 C, 35 - 85% RH |
Power Consumption | 5VA at 100 V AC |
Dimension mm H x W x D | 48 x 96 x 161 DIN |
Weight | Approx 600 gm |
Dielectric Strength | 500 V DC
between input/earth and case 500 V DC between input terminal/common, digital common, sensor power supply and relay output. 1500 VAC between power supply terminal/common, digital common, sensor power supply case and relay output |
Insulation Resistance | DC 500V 100 MOhm at above terminals |
Noise rejection | Power terminals, normal/common mode +/- 1500V square waveform for 1 ns at rise, noise, width 500 ns (by noise simulation) |
Sensor Power Supply | 10 V +/- 10%, 80 mA |
Optional Outputs |
BCD output (isolated from input (LO)) | |
Measured data | Negative
logic NPN open-collector outpug |
Polarity signal | Transistor on at minus input |
OVER signal | Transistor on at
"OVER" input |
Printing Command Signal | Transistor on for approx 10 ms at end of measurement |
ENABLE: | All data output
transistors arte turned "OFF" when "ENABLE" and digital common terminals are
shorted or input level is set at "0". "0" level less than 1.5 V "1" level 3 to 30 V input current < -5 mA |
Transistor output capacity | Applied Voltage
30 V max, Current 10mA max. Output saturation voltage less than 1.2 V at 10 mA TTL output is also available |
RS232C (isolated from input (LO)) | |
Communication method | Full duplex |
Synchronous method | Start and Stop |
Transmission speed | 2400/4800/9600/19200 bps |
Start Bit | 1 Bit |
Data length | 7 Bit |
Error detection | Even parity |
Stop Bit | 2 Bits |
Character Code | ASC II |
Transmission control | none |
RS485 (isolated from input (LO)) | ||||||||||
Communication method | 2 wire system half duplex | |||||||||
Synchronous method | Start and Stop | |||||||||
Transmission speed | 2400/4800/9600/19200 bps | |||||||||
Start Bit | 1 Bit | |||||||||
Data length | 7 Bit | |||||||||
Error detection | Even parity BCC (Block Check Character) Check sum |
|||||||||
Stop Bit | 2 Bits | |||||||||
Character Code | ASC II | |||||||||
Transmission control | none | |||||||||
|
||||||||||
Number of connectable meters | Up to 31 units | |||||||||
cable length | up to 500 m in total |
Analog Output (isolated from input
(LO)) Any display range to output analog signal can be set |
|||||||||||||
Resolution | 16 Bits | ||||||||||||
Output Response | < 0.5 secs | ||||||||||||
|
|||||||||||||
Moving Average | Digital Filter Function | ||||||||||||
Fixed zero | 10e0 digit is fixed to "0" for display | ||||||||||||
Display Blanking | Blanking of LEDs on main, monitor/sub display units. |
![]() |
Click for larger image |
Ordering Information |
Topac Inc. 231 CJC Highway, Suite 103
Cohasset MA 02025 USA
Tel: 781 740 8778 Fax: 781 740 8779
e-mail:
sales@topac.com
| Home | E-mail | Online store | Inquiry form | Related |
This is a world wide web document page design
layout and HTML code copyright ©1999 by Topac Inc. - ALL RIGHTS
RESERVED